5.6.4 ChemoClean function
Just like the timer function, ChemoClean can
also be used to start a cleaning cycle.
However, ChemoClean supports different
cleaning and rinse intervals.
Thus, irregular cleaning with different repeat
cycles is possible, and cleaning times with
post-rinse times can be individually defined.
Note:
• Use relays 3 (water) and
4 (cleaner) for the ChemoClean
function.
• Abortion of the cleaning process is
always followed by a post-rinse
time.
• When “Economy” is selected,
cleaning is performed with water
only.
5.5.5 Neutralisation controller
Note:
Use relays 1 and 2 for the
neutralisation control.
Neutralisation control means that the pH value
of a medium is held constant by adding acid
or alkali as required. This task requires two
separate actuating signals – one for acid and
one for alkali.
The neutralisation controller is a controller
with two relay contacts specifically tailored to
this application. A P(ID) controller is available
to handle this task.
The controller gain K
p
for acid and alkali can
be separately adjusted. Integral action time T
n
and derivative action time T
v
apply to both
controllers (compared to chap. 5.5.2, p. 39).
The “neutral zone” is located between the set-
point 1 and setpoint 2. In the case of a control-
ler without an integral component (P, PD),
there is no alkali or acid dosing ink the “neu-
tral zone” (Y = 0, see Fig. 5.16). In the case of
a controller with an integral component (PI,
PID), there is constant alkali/acid dosing (Y
new
=Y
old
). The behaviour of the I component
within the “neutral zone” depends on the
process type (Inline / Batch).
The “neutral zone” can be shifted in the X
direction as desired via the set points 1 and 2.
Hold
Water
Cleaner
CHEMO.CDR
t
0
t
1
t
2
t
3
t
4
t
5
t
1
cleaning start
t
2
– t
1
pre-rinse time
t
3
– t
2
cleaning time
t
4
– t
3
post-rinse time
t
5
– t
4
hold dwell period
Fig. 5.15 Cleaning cycle sequence
NEUTRA.CDR
Y [%]
Regulating contact 1
for alkali
Regulating contact 2
for acid
neutral
zone
X
100
50
0
Set point
Setpoint 1
Setpoint 2
Fig. 5.16
Characteristic of a
proportionalneutralisation
controller
Instrument configuration Liquisys M CPM 223 / 253
42 Endress+Hauser