7.2 Troubleshooting using the error messages
Display and select error messages by pressing the MINUS key.
Error
no.
Display Measures
Contakt Error current
Automatic start
of cleaning
Fact User Fact User Fact User
E001
EEPROM memory error
Switch instrument off and back on,
return instrument to your local
Endress+Hauser sales agency for repair
or replace instrument.
Load software compatible with hardware.
Load measuring parameter specific
instrument software.
yes no ——*
E002
Instrument not calibrated,
calibration data invalid,
no user data invalid
(EEPROM error.
Instrument software not
suitable for hardware
(controller).
yes no ——*
E003
Download error
Invalid configuration. Repeat download,
check optoscope.
yes no no
E004
Instrument software
version not compatible
with module hardware
version
Load software compatible with hardware.
Load measuring parameter specific
instrument software.
yes no no
E007
Transmitter malfunction
Software not compatible
with hardware (transmitter)
yes no ——*
E008
SCS alarm:
glass electrode: glass
breakage
IsFET: Leakage current
> 400nA
Check glass electrode for glass breakage
and hairline cracks; examine electrode
plug-in head for moisture and dry if
necessary; check medium temperature.
Replace IsFET.
yes no no
E010
Temperature sensor
defective, not connected
of short-circuited
Check temperature sensor and
connection; if necessary, check
instrument and measuring cable with
temperature simulation.
Check correct selection in field A5.
yes no no
E030
SCS reference electrode
error
Inspect reference code for soiling and
damage; clean reference electrode; check
medium temperature.
yes no no
E032
Slope range exceeded or
below range
Repeat calibration and renew buffer
solution; replace electrode if necessary,
and check instrument and measuring
cable with simulator.
yes no ——*
E033
pH value zero too low or
too high
yes no ——*
E034
Redox offset range
exceeded or below range
yes no ——*
E041
Calculation of calibration
parameters aborted
Repeat calibration and renew buffer
solution; replace electrode if necessary,
and check instrument and measuring
cable with simulator.
yes no ——*
E042
Distance of calibration
value for buffer pH2 from
zero (pH7) too short
Slope calibration requires a buffer solution
with a difference of at least pH = 2 from
the electrode zero point.
yes no ——*
E043
Distance between
calibration values for pH1
and pH2 too short
Use buffer solutions that are at least pH
= 2 apart.
yes no ——*
PM253E07.CHP
Liquisys M CPM 223 / 253 Maintenance and troubleshooting
Endress+Hauser 63