Error
Possible cause Remedy Equipment need, spare parts
Fixed, incorrect
measure value
Sensor not immersed or
protective cap not removed
Check installation, remove
protective cap
Air cushion in assembly Check assembly and installation
Short to ground on/in instrument Test measurement in insulated
container, poss. with buffer solution
Plastic container, buffer solutions
Hair crack in glass membrane Replace sensor pH sensor
Impermissible instrument
operating state (no response to
key actuation)
Switch instrument off and back on EMV problem: check grounding
and line routing if problem persists
Incorrect temperature
value
Incorrect sensor connection Verify connections using
connection diagram
Connection diagram chap. 3.4
Measuring cable defective Check cable Ohmmeter
Incorrect sensor type Select sensor type on instrument
(field B1)
Glass electrode: Pt 100
IsFET: Pt 1000
Incorrect pH value in
process
No / incorrect temperature
compensation
ATC: activate function
MTC: set process temperature
Conductivity of medium too low Choose pH sensor with salt
reservoir or liquid KCI
e.g. Orbisint CPS 11-xASxxxx
e.g. Ceraliquid CPS 41
Flow rate too high Reduce flow rate or measure in a
bypass
Potential in medium Try grounding with/at PA pin
(PA/PE connection)
Problem mainly occurs with plastic
lines
Sensor soiled or coated Clean sensor (see chap. 8.6.2) Heavily soiled media: use spray
cleaning
Measured value
fluctuates
Measuring cable interference Connect cable screen acc. to
connection diagram
See chap. 3.5
Signal output line interference Check line routing, try separate
line routing
Signal output and meas. input lines
Interference potential in medium Use symmetric measurement (with
PMC)
Try grounding medium via PA/PE
connection
No potential matching (PA/PM)
with symmetrical input
Connect PA pin in assembly to
PA/PM on instrument
Controller or timer
cannot be activated
No relay module installed Install LSR1-2 or LSR1-4 module See chap. 8.2 and 8.3
Controller / limit contact
does not work
Controller switched off Activate controller See chap. 5.5 or fields R2xx
Controller in Manual / Off mode Choose Autoor Manual On
mode
Keyboard, REL key
Pickup delay setting too long Disable or shorten pickup delay See fields R2xx
–“Hold function active Autom. hold during calibration,
hold-input activated;
holdvia keyboard active
See fields S2 to S4
Controller / limit contact
works continuously
Controller in Manual / Onmode Set controller to Manual / Offor
Auto
Keyboard, REL and AUTO keys
Dropout delay setting to long Shorten dropout delay See fields R2xx
Control loop interruption Check measured value, current
output or relay contacts, actuators,
chemical supply
No pH/mV current
output signal
Line open or short-circuit Disconnect line and measure
directly in instrument
mA meter 020 mA DC
Output defective See chap. 8.1
Fixed current output
signal
Currenct simulation active Switch-off simulation See field O2
Processor system out of sync Switch instrument off and back on EMV problem: Check installation if
problem persists
Incorrect current output
signal
Incorrect current assignment Check current assignment:
020 mA or 420 mA?
Field O211
Total load in current loop
excessive (> 500 )
Disconnect output and measure
directly on instrument
mA meter for 020 mA DC
Current output table not
accepted
Value interval too small Use sensible intervals
PM253E07.CHP
Liquisys M CPM 223 / 253 Maintenance and troubleshooting
Endress+Hauser 61