Error
Possible cause Remedy Equipment need, spare parts
No temperature output
signal
– Instrument only has one current
output
Refer to nameplate for variant,
change LSCH-x1 module if nec.
LSCH-x2 module,
see chap. 8.2.4 and 8.3.4
– Instrument with Profibus PA PA instrument has no current output!
Chemoclean function
not available
– No relay module (LSR1-x) or only
LSR1-2 installed
Install LSR1-4 module.
Chemoclean is enabled via release
code supplied by E+H with
Chemoclean upgrade
LSR1-4 module,
see chap. 8.2.4 and 8.3.4
Plus ackage functions
not available
– Plus package not enabled
(enable with code that depends
on serial number and is received
from E+H with order of Plus
package)
– Plus package upgrade:
code received from E+H ⇒ enter
– Following replacement of
defective LSCH/LSCP module:
first enter instrument serial
number (see nameplate)
manually, then enter code
Detailed description
see chap. 8.3.5
No HART or PROFIBUS
communication
– Several devices on the same
address
Check addresses and reset if
necessary
No HART
®
-
communication
– No central HART module Verify by looking at nameplate:
HART = -xxx5xx und -xxx6xx
Upgrade to LSCH-H1 / -H2 module
– No or wrong DD
(device description)
For further information, see
additional Operating Instructions
BA 208C, “HART
®
– Field
communication with Liquisys M
CxM 223 / 253”
– HART interface missing
– Instrument not registered with
HART server
– Load < 230 Ω
– HART receiver (e.g. FXA 191) not
connected via load
– Incorrect device address
(addr. = 0 for single operation,
addr. > 0 for multidrop operation)
– Line capacitance too high
– Line interference
No PROFIBUS
®
-
communication
– No central PA module Verify by looking at nameplate:
PROFIBUS-PA = -xxx3xx
Upgrade to LSCP module
– Wrong SW version (without PA)
For further information, see
additional Operating Instructions
BA 209C, “PROFIBUS-PA/-DP –
Field communication with
Liquisys M CxM 223 / 253”
Notes on project planning for
PROFIBUS
®
can be found in
Technical Information TI 260F,
detailed information on
instrumentation and accessories in
the operating instructions BA 198F.
– Commuwin (CW) II:
CW II version and instrument
SW version incompatible
– No or wrong DD/DLL
– Incorrect baud rate setting for
segment coupler in DPV-1 server
– Incorrect station (master)
address or duplicate address
– Incorrect station (slaves) address
– Bus line not terminated
– Line problems
(too long, cross section too
small, not shielded, screen not
grounded, wires not twisted)
– Bus voltage too low (bus supply
voltage typ. 24 V DC for non-Ex,
13,5 V DC for Ex)
Voltage at instrument PA connector
must be at least 9 V.
Maintenance and troubleshooting Liquisys M CPM 223 / 253
62 Endress+Hauser