Error
Possible cause Remedy Equipment need, spare parts
No temperature output
signal
Instrument only has one current
output
Refer to nameplate for variant,
change LSCH-x1 module if nec.
LSCH-x2 module,
see chap. 8.2.4 and 8.3.4
Instrument with Profibus PA PA instrument has no current output!
Chemoclean function
not available
No relay module (LSR1-x) or only
LSR1-2 installed
Install LSR1-4 module.
Chemoclean is enabled via release
code supplied by E+H with
Chemoclean upgrade
LSR1-4 module,
see chap. 8.2.4 and 8.3.4
Plus ackage functions
not available
Plus package not enabled
(enable with code that depends
on serial number and is received
from E+H with order of Plus
package)
Plus package upgrade:
code received from E+H enter
Following replacement of
defective LSCH/LSCP module:
first enter instrument serial
number (see nameplate)
manually, then enter code
Detailed description
see chap. 8.3.5
No HART or PROFIBUS
communication
Several devices on the same
address
Check addresses and reset if
necessary
No HART
®
-
communication
No central HART module Verify by looking at nameplate:
HART = -xxx5xx und -xxx6xx
Upgrade to LSCH-H1 / -H2 module
No or wrong DD
(device description)
For further information, see
additional Operating Instructions
BA 208C, HART
®
Field
communication with Liquisys M
CxM 223 / 253
HART interface missing
Instrument not registered with
HART server
Load < 230
HART receiver (e.g. FXA 191) not
connected via load
Incorrect device address
(addr. = 0 for single operation,
addr. > 0 for multidrop operation)
Line capacitance too high
Line interference
No PROFIBUS
®
-
communication
No central PA module Verify by looking at nameplate:
PROFIBUS-PA = -xxx3xx
Upgrade to LSCP module
Wrong SW version (without PA)
For further information, see
additional Operating Instructions
BA 209C, PROFIBUS-PA/-DP
Field communication with
Liquisys M CxM 223 / 253
Notes on project planning for
PROFIBUS
®
can be found in
Technical Information TI 260F,
detailed information on
instrumentation and accessories in
the operating instructions BA 198F.
Commuwin (CW) II:
CW II version and instrument
SW version incompatible
No or wrong DD/DLL
Incorrect baud rate setting for
segment coupler in DPV-1 server
Incorrect station (master)
address or duplicate address
Incorrect station (slaves) address
Bus line not terminated
Line problems
(too long, cross section too
small, not shielded, screen not
grounded, wires not twisted)
Bus voltage too low (bus supply
voltage typ. 24 V DC for non-Ex,
13,5 V DC for Ex)
Voltage at instrument PA connector
must be at least 9 V.
Maintenance and troubleshooting Liquisys M CPM 223 / 253
62 Endress+Hauser