Error
no.
Display Measures
Contakt Error current
Automatic start
of cleaning
Fact User Fact User Fact User
E154
Below lower alarm
threshold for period
exceeding alarm delay
If necessary perform manual reference
measurement. Repair sensor and
recalibrate.
yes no no
E155
Above upper alarm
threshold for period
exceeding alarm delay
yes no no
E156
Actual value undershoots
alarm threshold for longer
than the set permissible
maximum period
yes no no
E157
Actual value exceeds
alarm threshold for longer
than the set permissible
maximum period
yes no no
E162
Dosage stop
Check settings in CURRENT INPUT or
CHECK function group.
yes no no
E164
Dynamic range of pH
convertor exceeded
Check cable and sensor. yes no -
E165
Dynamic range of
reference convertor
exceeded
Check cable and sensor. yes no -
E168
Warning: IsFET leakage
current > 200 nA
Check IsFET for abrasion and tightness,
replace soon.
no no no
E171
Flow in main stream loo
low or zero
Restore flow. yes no no
E172
Switch-off limit for current
iput exceeded
Check process variables at sending
measuring instrument.
yes no no
E173
Current input < 4 mA
Check process variables at sending
measuring instrument.
Change range assignment if required.
yes no no
E174
Current input > 20 mA
Check process variables at sending
measuring instrument.
Change range assignment if required.
yes no no
PM253E07.CHP
Liquisys M CPM 223 / 253 Maintenance and troubleshooting
Endress+Hauser 65